Hemo-Force® Go
Portable DVT pump for patients at home.
Many factors determine DVT risk, but studies show
the highest potential for a clot is 2–10 days after
surgery, with patients remaining at risk for up to
three months.
With hospital stays getting shorter,
risk is highest when patients are recovering at home.
• Helps prevent deep vein thrombosis (DVT) by promoting blood flow
• Reduces post-operative pain & swelling
• Comfortable, portable, easy-to-use compression therapy
• Cordless device—less risk for falls
• One size fits most
Hook & Loop Fasteners
Adjust quickly and securely over leg for a snug, custom fit.
The Cuff is made of Polyester lint, TC, sponge and PVC.
Characteristics and Advantages
• One-click simple operation
• Low pressure indicator
• Light weight
• Internal lithium battery operated
• Detachable to replace DVT sleeves
System Components
• Power Indicator
• Error Indicator
• Start/Stop Button
• Cuff
• USB Charging Port
• Cuff with tube connection port
• Single person use
• Soft and breathable cuff
• Compact design
• USB charger
Power Indicator Error Indicator Start/Stop
Cuff
Charging Port Tube Connection Port
Power Indicator
1. Green: When the unit power is ON and battery is charging, the green LED flashes; a steady green LED indicates battery is fully charged.
2. Red: When the device is in a powered-on state and the light is red flashing with a beep sound, it indicates that the battery level has dropped to the critical level and the device will automatically shut down after 2 minutes. Please turn it off for charging.
Error Indicator Solid yellow when error.
Powered by Internal Battery
The unit is powered by a built-in battery.
Charge for at least 11 hours before using for the first time or after using it for a long time.
[Normal Operation] If power indicator is “GREEN”, the unit is fully operational.
[Recharge] If the device does not turn on or appears red solid, it should be charged immediately.
[Charging] The device is turned off, the red/green light flashes alternately, when the battery is full, the red light is off, and the green light is solid.
[Full] Once the battery is full the green light does not flash, and it can be operational.
Product Specification
Power Adapter: Input: AC 100–240V, 50–60Hz; Output: DC 5V, 2A
Power Battery: Lithium battery 7.4V 650mAh Continuous working time (fully-charge): about 8 hours Charging time: about 4 hours
Air Bladder Quantity: 1 piece
Typical Pressure: 45 ±10% mmHg
Cycle Period: Inflation time 12 seconds, deflation time 48 seconds, cycle time 60 seconds.
Safety Protection: Automatic pressure relief when the pressure reaches 108mmHg.
Alarm Category: Low battery, battery exhaustion, incorrectly worn, pressure sensor failure. System Operating Environment:
2 years warranty
COLOR |
|
LATEX FREE |
|
LENGTH INCHES |
|
PRESSURE RANGE |
|
PRODUCT FUNCTION |
|
SIZE |
|
GTIN |
|